Analayzer Dissolved Inorganic Carbon with LI-7815
Gas analyzer: LI-7815; included; externally mounted
Carrier gas: CO2-free compressed air; 16 to 18 PSI from regulatory
Precision: ± 0.1% or better at DIC level ~ 2000 μmol/L
Accuracy: ± 0.1% or better at DIC level ~ 2000 μmol/L
Measurement time: ~ 4 minutes per measurement
Sample volume: 0.5 to 4 mL per measurement
Dimensions: 43 × 27 × 47 cm (H × W × D),
main unit only, not including LI-7815
Precision 0.1% nominally; up to 0.05%
Measurement Time 4 minutes per aliquot
Takes precise measurements across a wide range of values
Can extract 100% of CO2 from water samples in 2 minutes
Stationary on ships and laboratories
Suited for culturing and incubation measurements
Power supply: 100 to 240 VAC 50/60 Hz
Fuse: 2-Amp slow blow
Power: 250 W
Computer operating system: Windows® 10 (included)
Communication: RS-485 or RS-232, via Serial-to-USB adapter (included)
Operating environment: Indoor
Operating temperature: 15 to 28°C recommended.
Temperature fluctuations will affect the precision of the measurement.
Operating humidity: Up to 85%